Operating flow range | 750 to 3250 ml/min 5 to 500 ml/min with low flow adapter kit |
Constant flow compensation (inches water back pressure) | 3000 ml/min at 15 inches 2500 ml/min at 20 inches 2000 ml/min at 30 inches 1000 ml/min at 40 inches |
Volumetric flow | Held within ± 2% of the set flow from 1 to 3 liters and ± 4% below 1 liter |
Flow fault | If flow drops more than 5% , pump stops and holds historical data. After 5 minutes in fault, auto-restart is attempted every 5 minutes up to 10 times. |
Run Time | Approximately 10 hours for 2000 ml/min up to 30 inches back pressure. Dependent on sample media used. |
Flow accuracy | ± 3% of set-point after calibration to desired flow |
Atmospheric pressure accuracy | ± 0.3 inches Hg |
Timing accuracy | 1 minute/month at 25 C |
Operating temperature range | 32 to 113 F (0 to 45 C) |
Charging temperature range | 32 to 113 F (0 to 45 C) |
Storage temperature range | -4 to 113 F (-20 to 45 C) |
Altitude range | do not use pump above 7, 500 ft |
Battery pack | Removable battery pack with rechargeable NiCad battery, 4.8 V x 2.0 Ah capacity |
Size | 5.6 x 3 x 2.3 inches (14.2 x 7.6 x 5.8 centimeters) |
Weight | 22 ounces (624 grams) |
Intrinsically-safe | UL and cUL Listed (Model 210-2003); CENELEC approved IS to EEx ia IIB T4 (Model 210-2002EX) |
Case | Impact-resistant RFI/EMI- shielded CE marked |