HHPC6 + Series Technical Specifications |
Particle Size Range |
0.3 μm - 10 μm in size channels per ISO 14644-1 (FS 209E)
industry standard. (HHPC 2+: 0.5 μm - 5 μm.) |
Number of Channels |
HHPC 6+ : 6 Channels ( 0.3 μm - 10 μm.)
HHPC 3+ : 3 Channels ( 0.3 μm - 10 μm.)
HHPC 2+ : 2 Channels ( 0.5 μm - 5 μm.) |
Weight | Ultra-lightweight at 1.5 lbs for comfortable handheld use |
Flow Rate | 2.83 Ipm. Take ISO 14644-1 class 5 compliant 0.3 μm samples in a minute |
Communications | Access particle counting data direct to PC via Ethernet and USB cable,or use USB memory stick |
Display | Large 3.5” high resolution 320 x 240 color |
Enclosure | High impact polycarbonate ABS for exceptional durability |
Dimensions | 10.7” x 3.9” x 2.1” Contoured for convenient single handed operation. |
Battery Duration |
Over 10 hours with typical use model. Minimum 5.5 hours continuous sampling. |
Battery Charging | 3.5 hours |
Data Storage | 10,000 records for the most demanding applications |
Counting Efficiency | 50% @ 0.3 μm; 100% for particles > 0.45 μm (per ISO 21501) |
Zero Count Level | 1 count / 5 minute (per JIS B9921) |
Concentration Limits | 4,000,000 / ft3 @ 10% coincidence loss (HHPC 6+ and 3+) 2,000,000 / ft3 @ 10% coincidence loss (HHPC 2+) |
Count Modes | Raw counts, N/ft3, N/m3, N/L in Cumulative or Differential mode |
Alarms | User selected particle channels and limits |
Operating Environment | 50°F to 104°F (10°C to 40°C) / < 95% non-condensing |
Accessories Included | 1. Power supply, Universal 110-240V 2. Zero count inlet-filter and zero count filter adaptor 3. 6' USB cable 4. Cradle for ease of charging and convenient USB and Ethernet communications (HHPC 6+ only) |